X100-ZDT paper and board internal bond tester for Z-direction tensile strength.
A 1 kN X100-1 based Z-direction tester for internal bond strength of paper and board, supplied with a 1000 N load cell and PBA-T541 fixture package for TAPPI T541 and ISO 15754 workflows.




For mills and converters that need delamination risk measured in the Z direction.
ZDT is not a generic tensile test. It measures the force needed to separate paper or board through thickness, so fixture parallelism, tape bond, compression step, dwell and failure mode matter as much as the machine capacity.
- Internal bond strength of paperboard and selected papers where cohesive strength is lower than the tape bond.
- Board delamination checks for printing, folding, punching, converting and packaging performance.
- Supplier and process comparison across ply structure, coating, refining, bonding and moisture changes.
- Routine mill QA and R&D studies where maximum force at delamination must be traceable.
X100-ZDT configuration from the Testometric datasheet.
| Frame capacity | 1 kN | Machine type | Single-column, bench mounted |
|---|---|---|---|
| Included load cell | 1000 N, others available | Included fixture | PBA-T541 fixture with 5 test blocks, 1 inch or 25 mm square |
| Force accuracy | Better than +/-0.5% of reading down to 1/1000 of load cell capacity | Data rate at PC | 500 Hz through USB |
| Crosshead travel | 420 mm | Throat depth | 81 mm from force axis to column |
| Speed range | 0.001 to 1000 mm/min | Speed accuracy | +/-0.1% under stable conditions |
| Position resolution | 0.0001 mm | Frame stiffness | 5 kN/mm |
| Weight | 25 kg | Environment | -10 to +40 C, 10 to 90% RH non-condensing |
| Electrical supply | 230 V, 1 phase, 50/60 Hz; 115 V option | Power | 0.2 kW |
Built around the paper and board Z-direction methods.
TAPPI T541
Internal bond strength of paperboard by Z-direction tensile loading, using double-sided tape, compression between platens and normal separation.
ISO 15754
Paper and board method for determining tensile strength in the Z direction. It applies to paper and board, not corrugated fibreboard.
Failure mode control
Valid interpretation depends on cohesive paper/board failure rather than tape release, platen misalignment or edge damage.
Force verification
Force measurement can be verified to ISO 7500-1 or ASTM E4 where the lab quality procedure requires traceable calibration.
For ZDT, the common mistake is treating the fixture as a simple pull test. The method needs controlled sample area, tape choice, compression step, dwell, platen parallelism and failure inspection.
Details to confirm before ordering.
- Standard version, test area, block size, tape specification and sample conditioning requirement.
- Board grade, thickness range, expected bond strength and whether failure must be reported by ply or visual mode.
- Compression force/time, dwell time, separation speed and result units used by your mill or customer.
- Calibration requirement, reporting format and export fields for lab information systems or SPC.
Questions buyers ask before specifying this Testometric system.
Use these answers to confirm whether this model, fixture package and WinTest method fit your lab or production workflow before quotation.
What is Z-direction testing used for?
Which methods are relevant for X100-ZDT?
Why is a dedicated ZDT fixture important?
Compare adjacent models before final selection.
Many special purpose projects overlap. These related pages help check whether force range, sample envelope, fixture geometry or throughput points to a better fit.
X100-TT Tissue Testing Machine
A dedicated Testometric X100-TT system for dry and wet tensile testing of tissue, hygiene paper and related low-force paper products.
X100-COF Coefficient of Friction Tester
A Testometric COF tester for static and dynamic coefficient of friction testing of films, paper, board, coated sheets and packaging surfaces.
X100-FT Single Fibre Tensile Tester
A dedicated Testometric fibre tester for low-force tensile testing of individual fibres with pneumatic fibre grips and controlled methods.
Need Z-direction internal bond testing for paper or board?
Send board grade, test standard, test area, tape requirement, expected force range and reporting units.